Surgically implanted aortic valve bioprostheses deform after implantation: insights from computed tomography
2020
Objective: Little is known about the prevalence and degree of deformation of surgically implanted aortic biological valve prostheses (bio-sAVRs). We assessed bio-sAVR deformation using multidetector-row computed tomography (MDCT). Methods: Three imaging databases were searched for patients with MDCT performed after bio-sAVR implantation. Minimal and maximal valve ring diameters were obtained in systole and/or diastole, depending on the acquired cardiac phase(s). The eccentricity index (EI) was calculated as a measure of deformation as (1 − (minimal diameter/maximal diameter)) × 100%. EI of 10% non-circular. Indic