Retardance parameters measurements by employing a rotating polarizer-analyzer polarimeter

2019 
In this work, we proposed the usage of a rotating polarizer-analyzer polarimeter to retrieve retardance parameters of a sample by treating it as an elliptical retarder that can be characterized by three parameters: total retardance, fast axis orientation and an ellipticity related parameter. By employing the Mueller matrix approach, we developed the demodulation algorithm to retrieve each parameter with the association of the Fourier series analysis. We present experimental results considering dextrose concentrations diluted in distilled water scaling in a range from 30mg/dl to 45mg/dl. For comparison purposes, we employed the method of rotating an analyzer and compare the optical activity variation for both methods.