Observation of 67 keV x-rays with a scintillation detector using proportional-mode silicon avalanche photodiode
2016
We developed a scintillation X-ray detector using a proportional-mode silicon avalanche photodiode (Si-APD). We report a prototype detector using a lead-loaded plastic scintillator mounted on a proportional-mode Si-APD (active area size: 3 mm in diameter), which is operated at a low temperature. Using 67.41 keV X-rays, we could measure pulse-height spectra of scintillation light with a charge-sensitive preamplifier at 20, 0, and −35°C. Time spectra of the X-ray bunch structure were successfully recorded using a wideband and 60-dB-gain amplifier in hybrid-mode operation of the Photon Factory ring. We obtained a better time resolution of 0.51 ns (full width at half-maximum) for the single-bunch X-ray peak at −35°C. We were also able to observe a linear response of the scintillation pulses up to 8 Mcps for input photon rates up to 1.4 × 108 photons/s.