Analytical Performance Verification of a Molecular Diagnostic for Cytology-Indeterminate Thyroid Nodules
2012
Objective: Our objective was to verify the analytical performance of the Afirma gene expression classifier (GEC) in the classification of cytologically indeterminate thyroid nodule fine-needle aspirates (FNAs). Design: Analytical performance studies were designed to characterize the stability of RNA in FNAs during collection and shipment, analytical sensitivity as applied to input RNA concentration and malignant/benign FNA mixtures, analytical specificity (i.e. potentially interfering substances) as tested on blood and genomic DNA, and assay performance studies including intra-nodule, intraassay, inter-assay, and inter-laboratory reproducibility. Results: RNA content within FNAs preserved in FNAProtect is stable for up to 6 d at room temperature with no changes in RNA yield (P = 0.58) or quality (P = 0.56). FNA storage and shipping temperatures were found to have no significant effect on GEC scores (P = 0.55) or calls (100% concordance). Analytical sensitivity studies demonstrated tolerance to variation i...