Corrosion Study of Amorphous Mg 67 Zn 29 Ca 4 Alloy

2015 
The aim of this paper was investigating the corrosion resistance of Mg67Zn29Ca4 metallic glass and evaluating the ability of use studied amorphous alloy for medical applications as biodegradable implants. For comparative purposes the corrosion studies of the amorphous Mg67Zn29Ca4 samples were performed in a few simulated body fluids at 37°C. The immersion tests were performed in Ringer’s solution, PWE (multielectrolyte physiological fluid) and 3%NaCl water solution. The amount of evolved hydrogen[ml/cm2] and a corrosion rate Vcorr [g/(day·m2)] of studied Mg67Zn29Ca4 alloy were compared. Surface morphology of metallic glass after corrosion study was performed using the scanning electron microscopy.In aim determination dissolution rate and amount of evolved hydrogen was simulated the environment of human body fluids during immersion tests. The work presents structure characterization and corrosion properties analysis of Mg67Zn29Ca4 bulk metallic glasses in the form of plates. Samples structure was analyzed by means of X-ray diffraction. Fractographic study a cross section of the sample in plates form was performed using scanning electron microscopy