Performance of medical radiographic X-ray systems in Greece for the time period 1998–2004

2007 
Abstract This study presents the results of the on-site inspections performed by the Greek Atomic Energy Commission (GAEC) on conventional X-ray systems, both in public and private medical radiology departments. A part of the inspection concerns the assessment of important radiographic parameters obtained according to a specified quality control protocol and the comparison of the measured parameter values with the corresponding acceptance limits. A total number of 1011 radiographic systems were inspected by the GAEC during the period 1998–2004, with 63.4% of them being privately owned. Analysis of 8 different operational parameters is carried out providing information on the overall performance, as well as on each parameter of the inspected X-ray systems. Tube voltage reproducibility values show the highest percentage of acceptability (98.9%, 99.5% for private and public owned radiographic systems respectively), while linearity of radiation output for private systems (72.5%) and time accuracy for public ones (72.7%) show the worst results. The comparison of the results for the private sector to those of a similar study carried out during the period 1995–1997 indicates a substantial improvement in X-ray systems performance. Higher level of improvement shows exposure time accuracy (12.2% percentile increase) and linearity of radiation output (12.5% percentile increase). Nevertheless, the situation can be further optimized if maintenance and quality control of the radiographic systems are carried out on a more regular basis.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    11
    References
    6
    Citations
    NaN
    KQI
    []