Fate of polytetrafluoroethylene monocusp pulmonary valves in an animal model.

2003 
Creation of a competent pulmonary monocusp valve facilitates transition from pressure to volume overload following right ventricular outflow tract reconstruction. To determine intermediate-term results and performance of the different types of polytetrafluoroethylene membrane used to construct monocusp valves and transannular patches, 12 infant lambs underwent excision of the native pulmonary valve and insertion of a monocusp valve and transannular patch made from one of 4 types of membrane. Echocardiography was performed after 3, 6, 9, and 12 months, and cardiac catheterization was carried out prior to animal sacrifice at 6 (n = 4) or 12 (n = 8) months. There was no postoperative morbidity or mortality. On echocardiography, 6 valves were mobile (50%), 4 had diminished mobility (33%), and 2 were fixed (17%) prior to sacrifice. At catheterization, mild, moderate, and severe pulmonary regurgitation was observed in 4 valves each (33%), with no stenosis. Right ventricular outflow tract reconstruction with pol...