Efficacy of the Ejector Flow-meter. A Scavenging Device for Anaesthetic Gases

1985 
Measurements of air concentrations of nitrous oxide and halothane in the breathing zone of the anaesthetist and the operating-room nurse were carried out during inhalation anaesthesia with a Mapleson D system. Gas removal was performed from inside the breathing system at the same rate as that of the fresh gas inflow by means of an ejector flow-meter. The concentrations of nitrous oxide and halothane were maintained below the Danish Threshold Limit Values of 100 and 5 parts per million, respectively, by using this type of scavenging. When these anaesthetics were used simultaneously, the reduced Threshold Limit Values were not exceeded during endotracheal anaesthesia.