Abstract Advanced neutron detectors for experiments at new spallation sources will require greater counting rate capabilities than previously attainable. This necessitates careful design of both detector and readout electronics. As part of a new instrument for protein crystallography at LANSCE, we are constructing a detector whose concept was described previously (IEEE Trans. Nucl. Sci. NS-46 (1999) 1916). Here, we describe the signal processing circuit, which is well suited for 3 He detectors with a continuous interpolating readout. The circuit is based on standard charge preamplification, transmission of this signal over 20 meters or so, followed by sample and hold using a second order gated baseline restorer. This latter unit provides high rate capability without requiring pole-zero and tail cancellation circuits. There is also provision for gain-adjustment. The circuits are produced in surface mounted technology.