Investigation of dental cone-beam CT pixel data and a modified method for conversion to Hounsfield unit (HU)

2018 
Objectives:To investigate the relationship in dental cone-beam CT (CBCT) between the manufacturer-reported image pixel data and a modified conversion to CT number densities in Hounsfield unit (HU).Methods:A standardized CT phantom was imaged using typical clinical parameters on CBCT from three manufacturers (Carestream 9300®, Carestream Health, Rochester, NY; J Morita 3D Accutomo®, J. Morita Mfg. Corp., Kyoto, Japan; and Planmeca Promax 3D®, Planmeca Helsinki, OY, Finland). Reconstructed axial slices were evaluated using regions of interest to ascertain the mean pixel value in five materials in the phantom. The Digital Imaging and Communications in Medicine data were also evaluated to determine if raw pixel data had been adjusted during the image reconstruction. A modified version of the existing manual HU conversion technique was applied, and the resultant slope and y-intercept were used to scale the pixel values ultimately to HU for all images.Results:The DICOM header data show that a modified rescale y...