A new method for high-bandwidth servo control of the power ratio between two Raman beams for cold atom interferometer

2020 
Light shift produced by the AC Stark effect is one of the major factors limiting the accuracy and long-term stability of a cold atom interferometer. The first order light shift can be canceled by fixing the power ratio of the Raman beams at a specified value. We report here a new method to stabilize the power ratio of the two Raman lasers with ∼100 kHz locking bandwidth, suppressing the effect of the first order light shift. We first mixed the two Raman lasers (at different optical frequencies) with a reference beam and then used two Schottky diode detectors to extract the corresponding beat note signals for each beam, which are much easier to be manipulated and processed as they are in the microwave band. The stability of the power ratio is improved by three orders of magnitude from 5.84 × 10−3 to 3.51 × 10−6 at 1 s averaging time and reaches 1.59 × 10−7 at 10 000 s integrating time when the servo loop is engaged. This method can be used in other precise quantum measurement based on the stimulated Raman transition and can be applied to compact inertial sensors.