NeoBeat offers rapid newborn heart rate assessment.

2021 
BACKGROUND Heart rate (HR) is used to guide interventions during delivery room (DR) neonatal resuscitation. Dry electrode ECG (NeoBeat) may detect HR more rapidly than pulse oximetry (PO) and portable ECG, but real-time comparisons of these devices are lacking. DESIGN/METHODS PO, ECG and NeoBeat were placed sequentially on newborns in the DR. Time for device placement and time to accurate HR acquisition were noted. RESULTS DR resuscitations of 28 preterm/term infants were observed. The NeoBeat was placed faster (ie, 3 s) than PO (20 s, p=<0.0001) and ECG (16 s, p=<0.0001). Total time from initiation of device placement to HR acquisition was fastest with NeoBeat (13 s) versus ECG (42 s, p<0.0001) and PO (105 s, p<0.0001) (duration values=median). CONCLUSIONS These observations in a small cohort of relatively well neonates demonstrate that the NeoBeat is significantly faster to place and consistently acquires HR faster than PO and ECG.