A TL/OSL emission spectrometer extension of the Risø reader.
2002
The integration of a low cost, compact sized spectrometer with the Riso reader is described. The luminescence light emitted by the sample is transmitted by an optical fibre onto a fixed entrance slit of a spectrograph. The light is measured with a high sensitivity 2048-element CCD-linear array detector (Avantes PC2000) sensitive in the UV-VIS region. The CCD array has a low readout noise and a photon sensitivity of 86 photons per count. Examples are given of both thermally and optically stimulated 3-D spectra showing the luminescence spectra as a function of temperature and time respectively. Spectra from CaF 2 :Tm (TLD-300) chips could be distinguished from the background (2 SD) at a 0.7 Gy dose level.