The use of portable intraoperative computed tomography scanning for real-time image guidance: A pilot cadaver study

2008 
BackgroundThis study was performed to assess the feasibility of using intraoperative computed tomography (CT) to provide real-time updates to image guidance systems (IGSs) during surgery.MethodsThe xCAT ENT portable intraoperative CT scanner (Xoran Technologies, Ann Arbor, MI) was used to acquire scans before, midway, and at the end of six cadaver dissections during the Southern States Rhinology Course, Augusta, GA, in October 2006. These scans were used to recalibrate three different IGSs used during the dissection. Time measurements were recorded and dosimetry was obtained from the cornea, sphenoid sinus (near the optic chiasm), and from the operative field during acquisition of the images. IGS accuracy was determined at the skull base and lamina papyracea. Surgeons were interviewed on benefits of real-time updates to the IGS after completion of dissections.ResultsThe xCAT ENT scanner was compatible with all three IGS platforms. The average time to update the IGS was 13 minutes. Radiation doses to the c...