EVALUATION OF THE METAL ION RELEASE OF MODULAR HIP PROSTHESES

2018 
IntroductionDual modular hip prostheses were introduced to optimize the individual and intra-surgical adaptation of the implant design to the native anatomics und biomechanics of the hip. The downside of a modular implant design with an additional modular interface is the potential susceptibility to fretting, crevice corrosion and wear [1–2]. The purpose of this study was to characterize the metal ion release of a modular hip implant system with different modular junctions and material combinations in consideration of the corrosive physiological environment.MethodsOne design of a dual modular hip prosthesis (Ti6Al4V, Metha®, Aesculap AG, Germany) with a high offset neck adapter (CoCrMo, CCD-angle of 130°, neutral antetorsion) and a monobloc prosthesis (stem size 4) of the same implant type were used to characterize the metal ion release of modular and non-modular hip implants. Stems were embedded in PMMA with 10° adduction and 9° flexion according to ISO 7206-6 and assembled with ceramic (Biolox® delta) o...