Missed Hearing Loss in Tinnitus Patients with Normal Audiograms

2019 
ABSTRACT The prevalence of tinnitus is positively correlated with hearing loss, although, tinnitus can also present alongside clinically normal pure-tone thresholds. As standard pure tone audiograms (PTA) only sample at octave or inter-octaves, they potentially can miss lesions between the tested frequencies. Here we investigate if tinnitus patients with normal audiograms have hearing loss missed by standard PTA testing, referred as “missed hearing loss” in the paper. Hearing thresholds in 106 tinnitus patients who have a normal PTA were tested using fine frequency resolution (1/24 octave step) audiometry, referred as precision PTA (P-PTA), at ±1/3 octave band centered at their tinnitus frequencies. Tinnitus pitch, loudness and residual inhibition were evaluated based on the result of P-PTA. DPOAEs were also tested to evaluate the function of outer hair cells (OHC). Using the P-PTA test, we found that 49% (52 out of 106) of tinnitus patients with normal audiograms showed sharply notched hearing loss and most of the notches were at their tinnitus frequencies. Using a fine frequency step (1/24 octave) to assess tinnitus pitch, the successful matching rate increased to 84%, significantly higher than the rate measured in traditional method (51%, Fisher’s test, P