Automated quantification of mitral valve regurgitation based on normalized centerline velocity distribution
2001
Previous echocardiographic techniques for quantifying valvular regurgitation are limited by factors including uncertainties for orifice location and a hemispheric convergence assumption that often results in overand underestimation of flow rate and regurgitant orifice area. Using computational fluid dynamics simulations, these factors were eliminated, allowing a more accurate assessment of regurgitation. A model was developed to allow automated quantification of regurgitant orifice diameter based on the centerline velocity data available from color M-mode echocardiography. The model, validated using in vitro unsteady flow data, demonstrated improved accuracy for orifice diameter (y=0.95x+0.38, r=0.96) and volume (y=1.18x-4.72, r=0.93).
Keywords:
- Correction
- Source
- Cite
- Save
- Machine Reading By IdeaReader
7
References
0
Citations
NaN
KQI