Rib artifacts in electron beam tomography: incidence and severity without and with the cone beam reconstruction algorithm.

2001 
Electron beam tomography (EBT) may be compromised by rib artifacts. Two hundred forty-seven abdominal studies were performed without (Group A, n = = 222) or with (Group B, n = 25) the cone beam algorithm. One hundred eighty-six (83.8%) and nine (36%) studies of Groups A and B, respectively, displayed some level of artifact In Groups A and B, major, minor, and no artifacts were found in 115 (51.5%) and 0 (0%), 71 (32.3%) and 9 (36%), and 36 (16.2%) and 16 (64%) patients. respectively (p < 0.01). The cone beam algorithm improves EBT studies of the abdomen.