Repeatability of ultrasonic determination of gingival thickness

2007 
The aim of the present study was to assess the degree of disagreement of ultrasonic measurements of gingival thickness at different teeth. Gingival thickness was determined in 33 volunteers with plaque-induced gingivitis. Facial/buccal gingiva was measured at the level of the gingival sulcus depth. Measurements were repeated after 2 and 4 weeks. A repeated measures, two-level (occasion, subject), variance components model revealed a within-subject variance of 0.187 mm2 resulting in a repeatability coefficient of 1.20 mm. When modeling level 1 (occasion) variance as a function of tooth type, respective error terms were used for calculating 95% repeatability coefficients for different teeth. Unreliable measurements were largely confined to upper and lower second and third molars. Error terms were lowest (0.03–0.05) at upper canines and first premolars as well as lower anterior teeth and premolars, where repeatability coefficients of 0.5 to 0.6 mm could be estimated. It was concluded that performance of the device was best at certain tooth types with rather thin gingiva. The present resolution and rather high degree of disagreement may preclude, however, detection of minute increases in thickness in the micrometer range, which seem to occur during gingivitis.