Pre-Clinical Evaluation of an Adult Extracorporeal Carbon Dioxide Removal System with Active Mixing for Pediatric Respiratory Support:

2014 
The objective of this work was to conduct pre-clinical feasibility studies to determine if a highly efficient, active-mixing, adult extracorporeal carbon dioxide removal (ECCO 2 R) system can safely be translated to the pediatric population. The Hemolung Respiratory Assist System (RAS) was tested in vitro and in vivo to evaluate its performance for pediatric veno-venous applications. The Hemolung RAS operates at blood flows of 350-550 ml/min and utilizes an integrated pump-gas exchange cartridge with a membrane surface area of 0.59 m 2 as the only component of the extracorporeal circuit. Both acute and seven-day chronic in vivo tests were conducted in healthy juvenile sheep using a veno-venous cannulation strategy adapted to the in vivo model. The Hemolung RAS was found to have gas exchange and pumping capabilities relevant to patients weighing 3-25 kg. Seven-day animal studies in juvenile sheep demonstrated that veno-venous extracorporeal support could be used safely and effectively with no significant adverse reactions related to device operation.