Initial evidence of a 50% reduction of contrast media using digital variance angiography in endovascular carotid interventions

2020 
Abstract Purpose In previous clinical studies Digital Variance Angiography (DVA) provided higher signal-to-noise ratio (SNR) and better image quality than Digital Subtraction Angiography (DSA). Our aim was to investigate whether this quality reserve of DVA provides an opportunity for the reduction of iodinated contrast media (ICM) in carotid X-ray angiography (CXA). Method Our prospective study enrolled 26 patients (67.0 ± 8.1 years) undergoing carotid percutaneous transluminal angioplasty. The SNR of DSA and DVA image pairs obtained by a standard (100 %, 6 mL ICM) or a low-dose (50 %, 3 mL ICM) protocol were compared. Visual evaluation of all images was performed by five specialists using a 5-grade rating scale. The quality of DSA100 and DVA50 videos was also compared. Results DVA provided more than two-fold SNR, the median SNRDVA/SNRDSA ratio was 2.06 (100 %) and 2.25 (50 %). In the visual evaluation, the DVA100 score (3.73 ± 0.06) was significantly higher than the DSA100 score (3.52 ± 0.07, Wilcoxon p  Conclusions Our data show that DVA allows a substantial (50 %) ICM reduction in CXA without affecting the quality and diagnostic value of angiograms.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    25
    References
    1
    Citations
    NaN
    KQI
    []