Accuracy of determination of changes in cardiac output by transcutaneous continuous-wave doppler computer
1984
The value of a previously validated portable, continuous-wave Doppler computer was assessed for measuring changes in cardiac output (CO). Simultaneous thermodilution and Doppler CO values were measured in triplicate in 16 patients undergoing clinical intervention with vasodilator therapy. A continuous-wave Doppler transducer was placed in the suprasternal notch and directed toward the ascending aorta and angled until the maximal velocity signal was obtained. The correlation coefficient was 0.92 (standard error of the estimate [SEE] = 0.48 liter/min) at rest; with intervention it was 0.88 (SEE = 0.52 liter/min). Our data indicate that the Doppler computer technique, when used in selected patients, is reliable in detecting changes in CO after vasodilator therapy. It may be of value in clinical situations in which hemodynamic monitoring is impractical.