Calibration of an EJ309 liquid scintillator using an AmBe neutron source

2017 
Abstract As the light output function of the EJ309 liquid scintillator is nonlinear with respect to the energy of the recoil proton and it is difficult to be predicted, it need to be measured. In the present work, an EJ309 liquid scintillator, 5.1 cm in diameter and 5.1 cm in length, is calibrated with an AmBe neutron source using the time-of-flight method. A Monte Carlo simulation is performed to improve the accuracy of the measured light output function. The obtained light output function is presented and it is compared with existing measurements. The energy resolution of the EJ309 Liquid scintillator is also illustrated.