Do airport metal detectors interfere with implantable pacemakers or cardioverter-defibrillators?

2003 
Abstract Objectives The aim of this study was to determine whether airport metal detector gates (AMDGs) interfere with pacemakers (PMs) or implantable cardioverter-defibrillators (ICDs). Background It is currently unknown whether AMDGs interfere with implanted PMs or ICDs. Methods A total of 348 consecutive patients (200 PM and 148 ICD recipients) have been tested for the occurrence of electromagnetic interference (EMI) within the electromagnetic field of a worldwide-used airport metal detector. Results No interference, such as pacing or sensing abnormalities, was observed in any of the 200 PM and 148 ICD patients; also no reprogramming occurred. Conclusions In vivo testing of PM and ICD systems showed no EMI with a standard AMDG. Clinically relevant interactions with implanted PMs or ICDs seem unlikely.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    16
    References
    64
    Citations
    NaN