Effects of computed tomography contrast medium factors on contrast enhancement

2011 
The various nonionic iodinated contrast media used in contrast computed tomography (CT) studies differ in terms of their composition, characteristics, and iodine concentration (mgI/ml), as well as the volume injected (ml). Compared with ionic iodinated contrast media, nonionic iodinated contrast media are low-osmolar agents, with different agents having different osmotic pressures. Using a custom-made phantom incorporating a semipermeable membrane, the osmotic flow rate (HU/s) could easily be measured based on the observed increase in CT numbers, and the relationship between the osmotic pressure and the osmotic flow rate could be obtained (r2=0.84). In addition, taking the effects of patient size into consideration, the levels of contrast enhancement in the abdominal aorta (AA) and inferior vena cava (IVC) were compared among four types of CT contrast medium. The results showed differences in contrast enhancement in the IVC during the equilibrium phase depending on the type of contrast medium used. It was found that the factors responsible for the differences observed in enhancement in the IVC were the osmotic flow rate and the volume of the blood flow pathways in the circulatory system. It is therefore considered that the reproducibility of contrast enhancement is likely to be reduced in the examination of parenchymal organs, in which scanning must be performed during the equilibrium phase, even if the amount of iodine injected per unit body weight (mgI/kg) is maintained at a specified level.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    10
    References
    1
    Citations
    NaN
    KQI
    []