High-resolution PIXE with crystal spectrometer and position-sensitive proportional counter

1989 
Abstract A crystal spectrometer combined with a position-sensitive proportional counter was developed for analysing particle-induced X-rays. Three satellite lines (KL 0 , KL 1 , and KL 2 ) of Si K α X-rays produced by 1.5 MeV H + ion bombardment were well resolved. Energy resolutions of 3 and 30 eV are obtained, respectively, for Si K X-rays (1.74 keV) and Cr K α1 X-rays (5.414 keV). This method is expected to give a high-resolution PIXE of reasonable efficiency with low background or high detection limit, especially, for analysing light-element impurities in heavy-element matrix, and also to be useful for studies of ion-atom collisions, such as bremsstrahlung production and chemical effects in ion-atom collisions.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    3
    References
    9
    Citations
    NaN