Cyclic and Torsional Fatigue Resistance of Seven Rotary Systems

2021 
Introduction: This study aims to evaluate the cyclic and torsional fatigue resistance of seven rotary systems. Methods and Materials: A total of 140 instruments were tested, 20 of each system including Genius (GN) 25/0.04, TruShape (TS) 25/0.06, Logic (LOG) 25/0.06 taper, Vortex Blue (VB) 25/0.06, ProTaper Gold (PTG) 25/0.08, Hyflex CM (HCM) 25/0.06 and Hyflex EDM (EDM) 25/0.08 taper. Cyclic fatigue resistance testing was performed using an artificial stainless steel canal with a curvature (60° angle and 5-mm radius) located at 5 mm from the tip. The files (n=10) rotated until fracture, and time was recorded in seconds. The torsional test was evaluated according to ISO 3630-1. Data were analysed with one-way ANOVA and Tukey’s tests (α=5%). The fractured surface of the instruments were assessed using scanning electron microscopy to confirm the type of fracture. Results: The cyclic fatigue resistance value of EDM was significantly higher than those of all tested instruments (P 0.05). The torsional test showed that PTG 25/0.08 had the highest torsional strength value of all instruments tested followed by VB and EDM (P 0.05). In relation to angular deflection, the GN; TS; HCM, and EDM showed significantly higher values until fracture than the other groups (P 0.05). Conclusion: Our in vitro study EDM group had the highest cyclic fatigue resistance among all the tested instruments. For the torsional test, the PTG showed highest torsional strength and lowest angular deflection values.