NUMERICAL STUDY ON BLOOD FLOW CHARACTERISTICS IN A ARTERIOVENOUS GRAFT WITH DELIVERED DOSE DURING HEMODIALYSIS

2011 
Hemodialysis is essential for patients with end stage renal failure. It is important to improve the patency rate and to minimize occurrence of the stenosis. Also, the blood flow to the artificial kidney can affect the blood flow characteristics through arteriovenous graft. Thus, the delivered dose are important factors for analyzing hemodynamic characteristics during hemodialysis access. In this study, the numerical analysis was performed for the effect of the delivered dose during hemodialysis access on the blood flow through the graft. As a result, The adverse pressure gradient occurred in case of a larger delivered dose through a catheter than standard dose and the flow instability increased. Also the circulation flow appeared largely at anastomotic site of the vein when the delivered dose was exceeded about half blood flow of inlet blood flow.