Assessing the depth of isoflurane anaesthesia during cardiopulmonary bypass
2018
Introduction:Bispectral index (BIS) and monitoring of end-tidal concentration may be associated with a reduction in the incidence of awareness during volatile-based general anaesthesia. An analogue of end-tidal concentration during cardiopulmonary bypass (CPB) is measuring exhausted isoflurane concentration from the oxygenator as an estimate to blood and, so, brain concentration. The aim of this study was to determine the relationships between oxygenator exhaust and blood concentrations of isoflurane and the BIS score during CPB when administering isoflurane into the sweep gas supply to the oxygenator.Methods:Seventeen patients undergoing elective cardiac surgery using CPB and isoflurane with BIS monitoring were recruited in a single-centre university hospital. Isoflurane gas was delivered via a calibrated vaporiser at the beginning of anaesthetic induction. Radial arterial blood samples were collected after the initiation of CPB and before aortic cross-clamping, which were analysed for isoflurane by gas ...