Direct Dose Measurement On Patient During Percutaneous Coronary Intervention Procedures Using Radiophotoluminescence Glass Dosimeters.

2016 
: The purpose of this research was to measure accurate patient entrance skin dose and maximum skin absorbed dose (MSD) to prevent radiation skin injuries in percutaneous coronary interventions (PCIs). We directly measured the MSD on 50 PCIs by using multiple radiophotoluminescence glass dosimeters and a modified dosimetry gown. Also, we analysed the correlation between the MSD and indirect measurement parameters, such as fluoroscopic time (FT), dose-area product (DAP) and cumulative air kerma (C-AK). There were very strong correlations between MSD and FT, DAP and C-AK, with the correlation between MSD and C-AK being the strongest (r = 0.938). In conclusion, the regression lines using MSD as an outcome value (y) and C-AK as predictor variables (x) were y = 1.12x (R2 = 0.880). From the linear regression equation, MSD is estimated to be ~1.12 times that of C-AK in real time.