HYDROLOGIC TEST SYSTEM FOR FRACTURE FLOW STUDIES IN CRYSTALLINE ROCK

1984 
. A hydrologic test system has been designed to measure the intrinsic permeabilities of individual fractures in crystalline rock. This report contains the design schematics and specifications necessary to build and/or modify this test system for other similar field applications. This system is used to conduct constant pressure/declining flow rate and pressure pulse hydraulic tests. The system is composed of four distinct units: (1) The Packer System, (2) Injection System, (3) Collection System, and (4) Electronic Data Acquisition System. The apparatus is built in modules so it can be easily transported and reassembled. It is also designed to operate over a wide range of pressures (0–300 psig) and flow rates (0.2–1.0 gal/min). This system has proved extremely effective and versatile in its use at the Climax Facility, Nevada Test Site.