A Personal Glucose Meter for Label-Free and Washing-Free Biomolecular Detection

2018 
We developed a label-free and washing-free method for biomolecular detection using a personal glucose meter (PGM). ATP was selected as a model target, and cascade enzymatic reactions promoted by hexokinase and pyruvate kinase were adopted to link the amount of ATP to glucose that is detectable by a hand-held PGM. In principle, the presence of target ATP enables hexokinase to catalyze the conversion of glucose to glucose 6-phosphate by providing a phosphate group to glucose, and thus the amount of glucose is decreased in proportion to the amount of ATP. In addition, adenosine 5′-diphosphate (ADP), which is generated after hexokinase-catalyzed enzymatic reaction, is recovered to ATP by a pyruvate kinase enzyme. The regenerated ATP is again supplemented to catalyze multiple rounds of cascade enzymatic reactions, leading to signal amplification. As a result, the change of glucose amount that is inversely proportional to ATP amount is simply measured by a hand-held PGM. By employing this strategy, we successfu...