REDUCING PELVIS RADIOGRAPH EXPOSURE IN CHILDREN USING A DOSE SIMULATION X-RAY RESEARCH SOFTWARE.

2021 
Pelvis radiography is a frequent X-ray examination. The objective of our study was to determine the minimum dose to be delivered without reducing the quality. We included 60 children having a pelvis X-ray in four groups that were equally represented by weight ranges. A software simulated, for each radiograph, six additional simulated photonic noise images corresponding to 100, 80, 64, 50, 40 and 32% of the initial dose. The 360 radiographs were blindly scored by two radiologists using a semi-quantitative Likert scale. There was no significant difference in scoring between the reference radiograph and simulated radiographs at 80% of the dose in children between 0 and 15 kg and over 35 kg. Inter-observer reproducibility was moderate to very good. Pelvis X-ray doses might be reduced by 20% in children in our institution. Software that produces simulated X-ray with decreasing dose might be a useful tool for an optimization process.