USEFULNESS OF A NONMACHINERY BASED SYSTEM FOR THE REINFUSION OF CELL-FREE AND CONCENTRATED AUTOGENOUS ASCITIC FLUID
2008
: A nonmachinery based system using gravity dependent flow for the treatment and reinfusion of ascitic fluid was developed, and its usefulness was assessed. In a preliminary study using bovine plasma, samples with protein concentrations below 5.0 g/dl were found to be treatable with this system. Bovine plasma containing blood, prepared to 0.5% hematocrit and with a protein concentration of 3.0 g/dl, was also treatable. We conducted a clinical study of 1,799 treatment sessions (1,495 using a machinery based system and 304 using a nonmachinery based system) of 343 patients with ascites refractory to various treatments. The recovery ratio of protein from the original ascitic fluid was 96% using the nonmachinery based system and 77% with the machinery based system (p < 0.01). Of 253 continuous reinfusions of ascitic fluid using the nonmachinery based system, the original ascitic fluid at protein concentrations below 2.5 g/dl was treatable. Original ascitic fluid below a hematocrit of 0.7% (protein concentration, 1.4 g/dl) was also treatable. This new procedure was simple and time and labor saving; the high recovery ratio of protein also demonstrated the usefulness of the new system.