Safety results from the treatment of 109 cerebral aneurysms using the Woven EndoBridge technique: preliminary results in the United Kingdom

2018 
OBJECTIVEThe Woven Endobridge (WEB) device has been in clinical use for the treatment of brain aneurysms for the past 4 years. Observational studies to assess clinical outcome and related complications have been published. Clear evidence is required to better understand the safety profile of the WEB device. The authors here present a multicenter series that provides a detailed safety analysis focused on patient selection, procedural events, and technical issues of treated patients throughout the United Kingdom (UK).METHODSA nationwide password-protected database was set up to collect anonymous information across the UK (14 centers). Complications and clinical outcome were analyzed for the initial 109 patients (112 procedures). An independent root cause analysis classified the complications into groups (procedural, disease, device, ancillary device, and other). The modified Rankin Scale (mRS) was used as a marker of clinical outcome.RESULTSEach of the 109 patients had 1 aneurysm suitable for WEB treatment ...
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    27
    References
    27
    Citations
    NaN
    KQI
    []