Noninvasive measurement of pressure gradient across a coronary stenosis using phase contrast (PC)‐MRI: A feasibility study

2017 
Purpose To investigate the feasibility of blood pressure difference measurement, ΔP, across the coronary artery using phase contrast (PC)-MRI for potential noninvasive assessment of the functional significance of coronary artery stenosis. Methods Three-directional velocities in the coronary arteries acquired using 2D-PC-MRI were used with the Navier-Stokes equations to derive ΔP. Repeat phantom studies were performed to assess the reproducibility of flow velocity and ΔP. ΔP derived using PC-MRI (ΔPMR) and that obtained using pressure transducer (ΔPPT) were compared. Reproducibility of coronary flow velocity was assessed in healthy controls (n = 11). Patients with suspected coronary artery disease (n = 6) were studied to evaluate the feasibility of ΔPMR measurement across a coronary stenosis. Results Phantom: Good overall reproducibility of flow velocity and ΔP measurements and excellent correlation (ΔPMR vs ΔPPT) was observed: intraclass correlation (ICC) of 0.95(Vz), 0.72(Vx), 0.73(Vy), and 0.87(ΔPMR) and R2 = 0.94, respectively. Human: Good reproducibility of coronary flow velocity was observed: ICC of 0.94/0.95(Vz), 0.76/0.74(Vx), and 0.80/0.77(Vy) at cardiac phase 1/2. Significant (p = 0.025) increase in ΔPMR was observed in patients (6.40 ± 4.43 mmHg) versus controls (0.70 ± 0.57 mmHg). Conclusion ΔPMR in the coronary arteries is feasible. Upon further validation using the invasive measure, ΔPMR has the potential for noninvasive assessment of coronary artery stenosis. Magn Reson Med, 2016. © 2016 International Society for Magnetic Resonance in Medicine.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    42
    References
    8
    Citations
    NaN
    KQI
    []