Quantification of and correction for left ventricular systolic long-axis shortening by magnetic resonance tissue tagging and slice isolation.

1991 
BACKGROUNDMeasurement of regional left ventricular (LV) function is predicted on the ability to compare equivalent LV segments at different time points during the cardiac cycle. Standard techniques of short-axis acquisition in two-dimensional echocardiography, cine computed tomography, and standard magnetic resonance imaging (MRI) acquire images from a fixed plane and fail to compensate for through-plane motion. The shortening of the left ventricle along its long axis during systole results in planar images of two different levels of the ventricle, leading to error in any derived functional measurements. LV systolic long-axis motion was measured in 19 normal volunteers using MRI.METHODS AND RESULTSWith a selective radio frequency (RF) tissue-tagging technique, three short-axis planes were labeled at end diastole and standard spin-echo images were acquired at end systole in the two- and four-chamber orientations. Persistence of the tags through systole allowed visualization of the intersecting short-axis t...