Mechanical effect of contrast injection on coronary artery diameter during angiography

1994 
The authors performed an angiographic study in six healthy anaesthetized pigs to investigate possible mechanical effects of contrast injection on coronary artery pressure and diameter during the period of image acquisition. Angiograms were acquired at 30 f/sec for repeated injections of 8 ml of contrast material by hand and with a power injector at rates of 2, 4, and 6 ml/sec. A 1.5 F micro-tip pressure transducer was used to measure intracoronary pressure at 1, 2, and 4 cm downstream from the injection site for each injection rate. The results show a progressive increase in intracoronary pressure with increasing rate of injection (p >