Liquid chromatography amperometric detection of catechol, resorcinol, and hydroquinone with a copper‐based chemically modified electrode

1992 
A copper-based chemically modified electrode (CME) has been constructed and characterized for flow-through amperometric detection of catechol, resorcinol, and hydroquinone. Novel potential dependence of the detector response was first obtained for these analytes at the Cu CME, where negative peaks together with positive ones were observed in one definite chromatogram using amperometric detection. Its advantages in chromatographic applications were demonstrated. From these observations it is proposed that the detector response was governed by formation of copper complexes with the solutes. A dynamic linear range over two orders of magnitude was obtained, when operating the detector at +0.10 V vs. SCE, from which ng detection limits were achieved.