Detection and quantification of trace organic contaminants in water using the FT-IR-attenuated total reflectance technique.

2010 
The Fourier transform infrared-attenuated total reflectance (FT-IR-ATR) technique has been used to detect and quantify the following volatile organic compounds (VOCs) in water: 1,1,1,2-tetrachloroethane, 1,1,2,2-tetrachloroethane, styrene, and tetrachloroethylene, among which the first three compounds were investigated at parts per million levels for the first time. Enhancement of the detection was made by (1) coating the ATR crystal with a hydrophobic polymer membrane, (2) optimizing the flow rate of the sample solution, (3) varying polymer membrane thickness, and (4) increasing the number of reflection bounces within the ATR crystal. Our flow rate optimization confirmed a previous finding that turbulent flow is more favorable than laminar flow in detecting the VOCs in water. However, decreases of ATR signal intensity were observed at very high turbulency due to analytes flowing too quickly through and exiting the ATR cell to be adsorbed onto the polymer membrane. The optimal membrane thickness was found...
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    35
    References
    28
    Citations
    NaN
    KQI
    []