Fabrication of ultra-low radioactivity detector holders for Edelweiss-II

2004 
Abstract In its first stage, the EDELWEISS-II dark matter experiment will use 21×320 g ionization-heat bolometers with NTD thermal sensors. To improve the present sensitivity of EDELWEISS to WIMP interactions by a further factor 100, ultra-low radioactivity detector holders exclusively made of copper and Teflon have been designed. The new design is using the relative expansion coefficients of copper, Teflon and germanium to hopefully ensure a dissipation-free detector holding. In order to validate this new holder, we need to compare the vibration behavior of the old holder and the new one. The amplitude of the bolometer vibrations is measured using the variation of capacitance between electrodes sputtered on the Ge crystal and reference electrodes fixed above the detector holders. We present noise measurements and compare them to the equivalent measurements using the detector holders of the Edelweiss-I experiment, which gave very satisfying results.