Optimizing the Elemental Sensitivity and Focal Spot Size of a Monolithic Polycapillary Optic Using Micro-X-Ray Fluorescence

1998 
A commercial micro-X-ray fluorescence (MXRF) instrument with an aperture X-ray guide was used to compare elemental sensitivities and focal spot sizes with those obtained by focusing the source with a monolithic polycapillary optic retrofitted into the system. The capillary provided an intensity gain of 125 at 4 keV vs. using a pinhole beam collimator; however, this gain advantage declined with increasing analyte line energy as a result of the capillary being designed shorter than its optimal length to fit into the commercial instrument. A minimum capillary focal spot FWHM of 36 {micro}m was achieved, whereas the smallest pinhole aperture available of 50 {micro}m in diameter produced a focal spot width of 69 {micro}m FWHM. Hence, better MXRF lateral resolution could be obtained with the capillary with a simultaneous improvement in elemental sensitivity.