Proposal and Verification of Simultaneous Measurement Method for Three Thermoelectric Properties with Film-Type Thermocouple Probe

2017 
A new simultaneous measurement method for the measurement of the three thermoelectric properties with a film-type thermocouple probe was proposed. Seebeck coefficient was measured using the steady-state condi-tion of the differential method. The electrical resistivity was measured us-ing the four-probe method and the thermal diffusivity is measured using the periodic heating method. The effectiveness of the proposed method was verified using constantan as a reference material. After describing the effectiveness of the method, the measurement of three thermoelectric properties of Bi0.3Sb1.7Te3, which is a thermoelectric material, was per-formed.