Transabdominal Ultrasonography, Computed Tomography and Electronic Portal Imaging for 3-Dimensional Conformal Radiotherapy for Prostate Cancer

2007 
Purpose: To evaluate the feasibility and accuracy of daily B-mode acquisition and targeting ultrasound-based prostate localization (BATTM) and to compare it with computed tomography (CT) and electronic portal imaging (EPI) in 3-dimensional conformal radiotherapy (3-D CRT) for prostate cancer. Patients and Methods: Ten patients were treated with 3-D CRT (72 Gy/30 fractions, 2.4 Gy/fraction, equivalent to 80 Gy/40 fractions, for α/β ratio of 1.5 Gy) and daily BAT-based prostate localization. For the first 5 fractions, CT and EPI were also performed in order to compare organ-motion and set-up error, respectively. Results: 287 BAT-, 50 CTand 46 EPI-alignments were performed. The average BAT-determined misalignments in latero-lateral, antero-posterior and cranio-caudal directions were –0.9 mm ± 3.3 mm, 1.0 mm ± 4.0 mm and –0.9 mm ± 3.8 mm, respectively. The differences between BATand CT-determined organ-motion in latero-lateral, antero-posterior and cranio-caudal directions were 2.7 mm ± 1.9 mm, 3.9 ± 2.8 mm and 3.4 ± 3.0 mm, respectively. Weak correlation was found between BATand CT-determined misalignments in antero-posterior direction, while no correlation was observed in latero-lateral and cranio-caudal directions. The correlation was more significant when only data of good image-quality patients were analyzed (8 patients). Conclusion: BAT ensures the relative positions of target are the same during treatment and in treatment plan, however, the reliability of alignment is patient-dependent. The average BAT-determined misalignments were small, confirming the prevalence of random errors in 3-D CRT. Further study is warranted in order to establish the clinical value of BAT.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    35
    References
    40
    Citations
    NaN
    KQI
    []