Development of a Semi-automated Superimposing Image Verification System Using a Template Matching Algorithm in Radiotherapy
2005
To analyze shifts in the isocenter of images we developed a semi‑automated superimposing image‑ verification system that is capable of automatically quantifying shifts in the isocenter through image analy‑ sis with a personal computer (PC) . The accuracy and usefulness of this software were examined through a comparison of nine portal images with a simulation image and by comparing nine portal images with a DRR image using a human pelvic phantom. The difference between the known magnitude of shift and the mag‑ nitude of shift detected with this method was analyzed as detection error. When the portal images were compared with the simulation image the 95% confidence interval (95% CI) of detection errors (meanSD) was 0.570.36 mm (95% CI: 0.49–0.65 mm) . When the portal images were compared with the DRR image the respective figures were 0.680.38 mm (95% CI: 0.59–0.77 mm) . No significant difference was noted be‑ tween these two categories of comparison (N.S) . The absolute detection error (meanSD) in all directions was 0.340.34 mm for the comparison of portal images with the simulation image and 0.410.36 mm for the comparison of portal images with the DRR image. This system seems to be appropriate for verification of the treatment field by improving the accuracy of radiotherapy as a method of computer‑assisted land‑ mark recognition during image comparison. Summary
Keywords:
- Correction
- Source
- Cite
- Save
- Machine Reading By IdeaReader
15
References
1
Citations
NaN
KQI