Rationale and feasibility of transcatheter pulmonary valve implantation in small conduits with the Edwards Sapien valves.

2020 
BACKGROUND Conduit dilatation above 110% and TPVI in conduits <16 mm is not recommended. However, if we want to reach normal values for RVOT diameters and diminish reintervention rates, pushing these boundaries is essential. METHODS Analysis of subsequent patients who underwent TPVI with Edwards Sapien valves in conduits ≤16 mm between 2010 and 2020. RESULTS In n = 33 cases median age was 13 years (5-20 y) and median weight 47 kg (15-91 kg). Preexisting RVOT grafts were n = 28 Contegra® conduits and n = 5 homografts (12 mm n = 15; 14 mm n = 11; 16 mm n = 7). Implanted were the Sapien (n = 8), Sapien XT (n = 10) and Sapien 3 valve (n = 15) with 20 mm (n = 4), 23 mm (n = 19), 26 mm (n = 9) and 29 mm (n = 1). Mean minimal RVOT diameter after TPVI was 22,7 ± 2,3 mm (18-30 mm) which is 150% of the mean minimal RVOT diameter before TPVI (15,1 ± 4,3 mm). Covered stents were used in n = 10 cases. Contained conduit rupture occurred in n = 7 cases (21%). Residual RVOT gradients of 5,7 ± 4,9 mmHg (0-18 mmHg) showed adequate RV unloading. CONCLUSION TPVI could be performed successfully in all patients. Dilatation above 150% and a valve/conduit diameter ratio up to 2,4 were well tolerated. There was a considerable amount of conduit rupture but all were confined without further need for intervention or surgery.