Standardisation of 85Sr with digital anticoincidence counting and half-life determination of the 514 keV level of 85Rb
2019
Abstract The activity of a 85 Sr solution was determined by means of fully digital 4πβ(LS)-γ anticoincidence counting. The measurements were carried out in a custom-built, combined TDCR / 4πβ(LS)-γ coincidence system, utilising a commercially available CAEN N6751C digitizer. The analysis of the experimental data, collected in list-mode format, was performed off-line by using the SoftKAM computer code developed at PTB. The data were also used to determine the half-life of the 514 keV level of 85 Rb which was found to be (1020.2 ± 6.0) ns.