A dynamic calibration method for the pressure transducer

2016 
Pressure transducer is widely used in the field of industry. A calibrated pressure transducer can increase the performance of precision instruments in the closed mechanical relationship. Calibration is the key to ensure the pressure transducer with a high precision and dynamic characteristic. Unfortunately, the current calibration method can usually be used in the laboratory with a good condition and only one pressure transducer can be calibrated at each time. Therefore the calibration efficiency is hard to meet the requirement of modern industry with high efficiency. A dynamic and fast calibration technology with a calibration device and a corresponding data processing method is proposed in this paper. Firstly, the pressure transducer to be calibrated is placed in the small cavity chamber. The calibration process only contains a single loop. The outputs of each calibrated transducer are recorded automatically by the control terminal. Secondly, LabView programming is used for the information acquisition and data processing. The performance of the repeatability and nonlinear indicators can be figured out directly. At last the pressure transducers are calibrated simultaneously in the experiment to verify the suggested calibration technology. The experimental result shows this method can be used to calibrate the pressure transducer in the practical engineering measurement.