Evaluation of fully automated assays for the detection of Rubella IgM and IgG antibodies by the Elecsys® immunoassay system

2014 
Abstract Screening for acute rubella infection in pregnancy is an important element of antenatal care. This study compared the sensitivity, specificity and reproducibility of two new, fully automated Elecsys ® Rubella IgM and IgG immunoassays designed for the Elecsys 2010, Modular Analytics E170, COBAS e-411 and COBAS e-601 and e602 analytical platforms, with current assays using serum from patients with primary rubella infections, vaccinated patients, patients with potentially cross-reacting infections and on routine samples in clinical laboratories in France, Germany and Italy. Both assays showed good within-run and within-laboratory precision. A sensitivity of 79.8–96.0% was demonstrated for Elecsys IgM in primary, early acute infection, consistent with existing assays. In samples obtained from routine antenatal screening, the Elecsys Rubella IgM assay revealed high specificity (98.7–99.0%). A significantly ( p