Continuous-Display Four-Channel Electroencephalographic Monitoring in the Evaluation of Neonates With Paroxysmal Motor Events

2001 
The objective of this study was to determine the value of continuous-display four-channel electroencephalographic (EEG) monitoring to distinguish epileptic from nonepileptic paroxysmal motor events. Five full-term neonates with paroxysmal motor events were included in the study. Nurses were instructed to print contiguous pages of the continuous-display four-channel EEG monitoring and to activate the event marker on the simultaneously conducted continuous video-EEG telemetry unit during each paroxysmal motor event. The printouts from the continuous-display four-channel EEG monitoring were interpreted and compared with the corresponding segments of continuous video-EEG telemetry. Thirty paroxysmal motor events were captured. Sixteen paroxysmal motor events were epileptic, and 14 were nonepileptic. The interpretation of the printouts of the continuous four-channel EEG monitoring concurred with an independent interpretation of the corresponding video-EEG telemetry segments in all of the events. Continuous-dis...
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    3
    References
    5
    Citations
    NaN
    KQI
    []